Medical DeviceView official recall notice →
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedi
Brand / Firm
Product
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Units affected
12,156 US
Description
SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)