Medical DeviceView official recall notice →
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M0
Brand / Firm
Product
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
Reason for recall
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Units affected
508,151 units
Description
FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01